Scintillation photon detection and event selection in high resolution positron emission tomography

  • 1. Engineering Div., Lawrence Berkeley Lab., Univ. of California, Berkeley, CA 94720

Description

The scintillation photon detection and event selection subsystem for the high spatial resolution Donner 600-channel positron emission tomograph is described. The tomograph spatial resolution of better than 3.0 mm, FWHM, is obtained by using 600 closely-packed bismuth germanate 3 mm thick scintillation detectors. Each detector has its own 13.5 mm diameter photomultiplier and event selection channel, consisting of two fast pulse amplifiers, a charge integrator, a differential pulse height discriminator and precision time delay for event strobing. The event selection channel is capable of starting a timing cycle from the first photoelectric emitted from the photocathode. It is also capable of discriminating between true events and noise and of eliminating events occurring simultaneously in two adjacent channels. Adjustments of the event strobing time and two discriminator levels are performed by a computer

Additional details

Publishing Information

Journal Title
IEEE Trans. Nucl. Sci.
Journal Volume
NS-34
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
326-331
ISSN
0018-9499
CODEN
IETNA

Conference

Title
Nuclear science and nuclear power systems symposium.
Dates
29-31 Oct 1986.
Place
Washington, DC (USA).

Optional Information

Secondary number(s)
CONF-861007--.