Performance of the Harwell-type size-selective dust sampler
Creators
- 1. Japan Atomic Energy Research Inst., Tokai, Ibaraki. Tokai Research Establishment
Description
Accuracy for measuring respirable dusts of the Harwell-type size-selective dust sampler has been studied by numerical analysis using the experimental data of calibration test. The test was made with uranine methylene-blue particles of uniform particle sizes (aerodynamic diameter dsub(a) = 1--17 μm, geometric standard deviation sigmasub(g) = 1.07--1.16) ; the sampler was examined at air flow-rate 6--50 l/min. Performance of the sampler at flow rate 33 l/min agrees with the lung deposition at a tidal volume of 750 cm3, and that at 50 l/min approximately with the lung deposition at tidal volume 1450 cm3. The performances is hardly influenced by variation of the flow rate less than 10%. The accuracy of this sampler for respirable dusts at flow rate 33 l/min is within a factor of 2 for MMAD <= 10 μm. (auth.)
Availability note (English)
MF available from INIS under the Report Number.Files
8281551.pdf
Files
(426.7 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:6bd4d167a628acd69fc291cf95b95db6
|
426.7 kB | Preview Download |
Additional details
Publishing Information
- Imprint Pagination
- 8 p.
- Report number
- JAERI-M--6246
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 8281551
- Subject category
- S42: ENGINEERING;
- Descriptors DEI
- ACCURACY; AEROSOL MONITORING; AIR SAMPLERS; CALIBRATION; DISTRIBUTION; DUSTS; FILTERS; FLOW RATE; LUNGS; PARTICLE SIZE; PERFORMANCE TESTING; RADIATION MONITORING; RETENTION
- Descriptors DEC
- BODY; LABORATORY EQUIPMENT; MONITORING; ORGANS; RESPIRATORY SYSTEM; SAMPLERS; SIZE; TESTING