Published 1991
| Version v1
Journal article
Measurement of conversion coefficients for calibrating individual dosemeters with respect to the operational dose quantities on the PMMA slab phantom
Description
The backscatter and depth dose factors and the conversion coefficients between air kerma free-in-air and the dose equivalents in 7 mg.cm-2 and 1 g.cm-2 depths on the central axis of the PMMA slab phantom which is now preferred by the ICRU for calibrating individual dosemeters in terms of the operational dose quantities of ICRU report 39, were measured in parallel with ionisation chambers and TLD. The measurements were performed for three sets of differently filtered X rays with tube voltages between 15 and 400 kV, and for gamma radiation. The results obtained correspond well with those of comparable Monte Carlo calculations. (author)
Additional details
Additional titles
- Augmented title (English)
- Polymethyl methacrylate
Publishing Information
- Journal Title
- Radiation Protection Dosimetry
- Journal Volume
- 37
- Journal Issue
- 2
- Series
- Radiat. Prot. Dosim.
- Journal Page Range
- 79-84
- ISSN
- 0144-8420
- CODEN
- RPDOD
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 23019177
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BACKSCATTERING; CALIBRATION; DEPTH DOSE DISTRIBUTIONS; DOSE EQUIVALENTS; ENERGY DEPENDENCE; FILTRATION; GAMMA DOSIMETRY; METHACRYLATES; ORGANIC POLYMERS; PHANTOMS; X-RAY DOSIMETRY
- Descriptors DEC
- CARBOXYLIC ACID SALTS; DISTRIBUTION; DOSIMETRY; MOCKUP; ORGANIC COMPOUNDS; POLYMERS; RADIATION DOSE DISTRIBUTIONS; SCATTERING; SEPARATION PROCESSES; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS; STRUCTURAL MODELS