Evaluation of the individual tube current setting in electrocardiogram-gated cardiac computed tomography estimated from plain chest computed tomography using computed tomography automatic exposure control

  • 1. Tottori Univ. Hospital, Yonago, Tottori (Japan)

Description

The aim of this study was to estimate the tube current on a cardiac computed tomography (CT) from a plain chest CT using CT-automatic exposure control (CT-AEC), to obtain consistent image noise, and to optimize the scan tube current by individualizing the tube current. Sixty-five patients (Group A) underwent cardiac CT at fixed tube current. The mAs value for plain chest CT using CT-AEC (AEC value) and cardiac CT image noise were measured. The tube current needed to obtain the intended level of image noise in the cardiac CT was determined from their correlation. Another 65 patients (Group B) underwent cardiac CT with tube currents individually determined from the AEC value. Image noise was compared among Group A and B. Image noise of cardiac CT in Group B was 24.4±3.1 Hounsfield unit (HU) and was more uniform than in Group A (21.2±6.1 HU). The error with the desired image noise of 25 HU was lower in Group B (2.4%) than in Group A (15.2%). Individualized tube current selection based on AEC value thus provided consistent image noise and a scan tube current optimized for cardiac CT. (author)