Published January 11, 1996 | Version v1
Journal article

ISOLTRAP: a tandem Penning trap system for accurate on-line mass determination of short-lived isotopes

  • 1. Gesellschaft fuer Schwerionenforschung mbH, Darmstadt (Germany)
  • 2. Mainz Univ. (Germany). Inst. fuer Physik
  • 3. McGill Univ., Montreal, PQ (Canada). Foster Radiation Lab.

Description

The tandem Penning trap mass spectrometer ISOLTRAP has been set up at the on-line mass separator ISOLDE at CERN/Geneva for accurate mass measurements of short-lived nuclei with T1/2 ≥1 s. The mass measurement is performed via the determination of the cyclotron frequency of an ion in a magnetic field. The design of the spectrometer matches the particular requirements for on-line mass measurements on short-lived isotopes. With the ISOLTRAP spectrometer masses of more than 70 radioactive nuclei have so far been determined with resolving powers exceeding one million and an accuracy of typically 10-7. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
368
Journal Issue
3
Journal Page Range
p. 675-697.
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Collaborations
ISOLDE Collaboration.