Published 1986
| Version v1
Journal article
Calculations of the energy and angle dependence of response of a simplified design of photon personal dosemeter and its relationship with the recommended ICRU calibration quantity
Creators
- 1. National Radiological Protection Board, Chilton (UK)
Description
The results are presented and discussed of calculations of the response, as a function of energy and angle, of a simplified design of personal dosemeter to measure Hsub(p)(10) for photon radiation. The design consists of a tissue-equivalent detector in the form of a 1 mm thick disc of radius 2.5 or 5 mm, covered with a hemisphere of tissue-equivalent material of 1 cm radius. (author)
Additional details
Publishing Information
- Journal Title
- Radiat. Prot. Dosim.
- Journal Volume
- 16
- Journal Issue
- 4
- Series
- Radiat. Prot. Dosim.
- Journal Page Range
- 319-323
- ISSN
- 0144-8420
- CODEN
- RPDOD
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 18070285
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DOSE EQUIVALENTS; DOSEMETERS; ENERGY DEPENDENCE; GAMMA DOSIMETRY; INCIDENCE ANGLE; KEV RANGE 10-100; KEV RANGE 100-1000; MEV RANGE 01-10; PERFORMANCE; PERSONNEL DOSIMETRY; TISSUE-EQUIVALENT DETECTORS
- Descriptors DEC
- DOSIMETRY; ENERGY RANGE; KEV RANGE; MEASURING INSTRUMENTS; MEV RANGE; RADIATION DETECTORS