Published January 2015 | Version v1
Journal article

An X-ray image intensifier for microsecond time-resolved experiments

  • 1. Japan Synchrotron Radiation Research Institute, 1-1-1 Kouto, Sayo, Hyogo 679-5198 (Japan)

Description

Persistence of an X-ray image intensifier with a YAG (P46, Y3Al5O12:Ce3+) phosphor in the output window was examined using X-ray pulses from a storage ring and a high-speed CMOS camera. Because of the fast decay of the YAG fluorescence (60 ns), persistence of CsI:Na+ in the input window dominates the decay of intensity of the image intensifier. As reported before, persistence of CsI:Na+ had two major components when fitted with two exponential functions, a fast one around 600 ns and a slow one about 7 μs. In addition, it was found that a slower component, which is small but takes tens of microseconds to decay, also exists. Thus, this detector should be used with caution at a time resolution higher than about 50 μs when high accuracy of measurement is required

Availability note (English)

Available from http://dx.doi.org/10.1088/1748-0221/10/01/T01002

Additional details

Publishing Information

Journal Title
Journal of Instrumentation
Journal Volume
10
Journal Issue
01
Journal Page Range
p. T01002
ISSN