Use of a single particle intra-cavity laser particle spectrometer for measurements of HEPA filters and filter systems
Description
Current tests of HEPA filters and/or filter installations using DOP aerosols and conventional forward light-scatter photometers are limited to measuring protection factors of 104 to 105. In addition, forward light-scattering photometers have markedly decreased sensitivity to less than 0.3 μm particles and basically measure only a scattering signal which is not uniquely related to any given concentration or size distribution of scatterers. These limitations require that high efficiency systems, such as multiple stage HEPA filters, be evaluated one stage at a time, a procedure which is quite often impractical for many existing air cleaning systems and which may be in error. In order to obviate these difficulties, a single particle intracavity laser particle spectrometer has been used to measure protection factors of up to 2.4 x 108 for multiple HEPA systems and individual double ply HEPA filters. Because of the instrumental size resolution, protection factor as a function of particle size can be determined from .06 μm to 2.9 μm. The lack of background enables single counts to be statistically significant. Since coincidence errors occur at particle concentration greater than 106/liter, a known dilution must be introduced to measure challenge concentrations greater than this. The dilution measurement may be accomplished with the aid of a forward light-scattering photometer
Availability note (English)
MF available from INIS under the Report Number.
Files
Additional details
Publishing Information
- Imprint Pagination
- 14 p.
- Report number
- LA-UR--76-1537
Conference
- Title
- 14. US ERDA air cleaning conference.
- Dates
- 2 Aug 1976.
- Place
- Sun Valley, Idaho, United States of America (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 8283868
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- AEROSOLS; AIR FILTERS; EFFICIENCY; ENVIRONMENTAL ENGINEERING; FEASIBILITY STUDIES; PARTICLE SIZE; PERFORMANCE TESTING; PHOTOMETERS
- Descriptors DEC
- COLLOIDS; DISPERSIONS; FILTERS; MEASURING INSTRUMENTS; POLLUTION CONTROL EQUIPMENT; SIZE; SOLS; TESTING
Optional Information
- Notes
- Available from NTIS. $3.50.
- Secondary number(s)
- CONF-760822--12.