Published December 2004 | Version v1
Journal article

An artifact caused by using the inadequate acquisition parameter Tc-99m ECD brain perfusion SPECT

  • 1. College of Medicine, Chonbuk National Univ., Chonju (Korea, Republic of)

Description

Although artifacts in tomographic imaging can arise from a number of sources this case presents an artifact caused by using inadequate acquisition parameters which do not match the collimator with which a camera equipped in Tc-99m ECD brain SPECT. Whenever the collimator in tomographic studies is changed, we should take care to use suitable parameters of acquisition protocol according to the collimator with which a camera is fitted. Other ways to prevent an error of this type is to use a point source

Additional details

Publishing Information

Journal Title
Korean Journal of Nuclear Medicine
Journal Volume
38
Journal Issue
6
Series
9 refs, 3 figs
Journal Page Range
p. 540-542
ISSN