An artifact caused by using the inadequate acquisition parameter Tc-99m ECD brain perfusion SPECT
- 1. College of Medicine, Chonbuk National Univ., Chonju (Korea, Republic of)
Description
Although artifacts in tomographic imaging can arise from a number of sources this case presents an artifact caused by using inadequate acquisition parameters which do not match the collimator with which a camera equipped in Tc-99m ECD brain SPECT. Whenever the collimator in tomographic studies is changed, we should take care to use suitable parameters of acquisition protocol according to the collimator with which a camera is fitted. Other ways to prevent an error of this type is to use a point source
Additional details
Publishing Information
- Journal Title
- Korean Journal of Nuclear Medicine
- Journal Volume
- 38
- Journal Issue
- 6
- Series
- 9 refs, 3 figs
- Journal Page Range
- p. 540-542
- ISSN
- 1225-6714
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 36036339
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BRAIN; CAMERAS; COLLIMATORS; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; TECHNETIUM 99; TOMOGRAPHY
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; CENTRAL NERVOUS SYSTEM; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; NERVOUS SYSTEM; NUCLEI; ODD-EVEN NUCLEI; ORGANS; RADIOISOTOPES; TECHNETIUM ISOTOPES; TOMOGRAPHY; YEARS LIVING RADIOISOTOPES