Published 1978
| Version v1
Book
Exposure intercomparison with ionization chambers. Based on three intercompared check sources
- 1. Instituto de Radioprotecao e Dosimetria, Rio de Janeiro (Brazil)
- 2. Gesellschaft fuer Strahlen- und Umweltforschung m.b.H., Neuherberg (Germany, F.R.). Inst. fuer Strahlenschutz
- 3. Physikalisch-Technische Werkstaetten Dr. Pychlau K.G., Freiburg im Breisgau (Germany, F.R.)
Description
The calibration factor of an ionization chamber can be related to the check reading caused by one or more overall radioactive stability check devices. It will be shown how one ionization chamber can be used for the intercomparison of three SSDLs. Each SSDL has its own radioactive check device which, for one given chamber, is used at each SSDL to normalize the calibration factor on the basis of a check reading. To carry out an intercomparison the chamber alone, i.e. without a radioactive source, can be sent from place to place. This procedure has great advantages from the customs and transport safety regulations point of view. (author)
Additional details
Publishing Information
- Publisher
- IAEA.
- Imprint Place
- Vienna
- ISBN
- 92-0-010478-9
- Imprint Title
- National and international standardization of radiation dosimetry
- Series
- Proceedings series.
- Journal Page Range
- v. 1 p. 213-218.
Conference
- Title
- International symposium on national and international standardization of radiation dosimetry.
- Dates
- 5 - 9 Dec 1977.
- Place
- Atlanta, GA, USA.
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 10438206
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; COMPARATIVE EVALUATIONS; IONIZATION CHAMBERS
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTORS
Optional Information
- Secondary number(s)
- IAEA-SM--222/36.