Published 1980 | Version v1
Miscellaneous

Experimental determination of dose distributions from Cf252 therapeutic sources

Description

A simple methods of evaluating dose distributions of therapeutic sources on the base of 252Cf, is presented. Radiophotoluminescent alumophosphate glasses of FYD-6 dosemeter of TOSHIBA firm (Japan) have been used as detectors. Dose, produced by γ-radiation of 252Cf, has been measured directly, and neutron radiation has been registered by measuring β-radiation, formed during 31Si decay, which is formed by 31P(n, p)31Si reaction during glass irradiation with fast neutrons β-radiation of 31Si has been registered by Vavilov-Cherenkov radiation, occurring in glass. Thus, it is possible to measure the dose of neutron and γ-radiation of sources simultaneously, disregarding contribution of thermal neutrons (<5%). Methods and applied devices are described, diagrams of radiation weakening are presented. Measuring error is determined. The suggested method is advisable in clinic investigations

Additional details

Additional titles

Original title (Russian)
Ehksperimental'noe opredelenie doznykh raspredelenij ot terapevticheskikh istochnikov iz

Publishing Information

Imprint Title
Radiation techniques. Issue 19
Journal Page Range
p. 200-204.
Report number
INIS-SU--57

Optional Information

Notes
6 refs.; 3 figs. Imprint:Radiatsionnaya tekhnika. Vypusk 19.;Voprosy atomnoj nauki i tekhniki.