Published 1996 | Version v1
Journal article

Differential pulse voltammetric determination of uranium in low concentration streams

  • 1. Bhabha Atomic Research Centre, Bombay (India)
  • 2. Bhabha Atomic Research Centre, Bombay (India). Analytical Chemistry Div.

Description

A differential pulse voltammetric method has been successfully used for the determination of uranium in low concentration streams of a uranium plant. The method gives a precision of about 13% to 7% in the range of 300 ppb to 15 ppm. The accuracy of the results was ascertained by comparing the values with those obtained by a spectrophotometric method. The method is simple, fast, sensitive, fairly accurate and does not require a preconcentration step. (author). 6 refs., 4 figs., 1 tab

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
207
Journal Issue
1
Journal Page Range
p. 39-43.
ISSN
0236-5731
CODEN
JRNCDM

INIS

Country of Publication
Hungary
Country of Input or Organization
Hungary
INIS RN
27073924
Subject category
S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
Descriptors DEI
ACCURACY; COMPARATIVE EVALUATIONS; DIAGRAMS; QUANTITATIVE CHEMICAL ANALYSIS; SAMPLE PREPARATION; URANIUM; VOLTAMETRY
Descriptors DEC
ACTINIDES; CHEMICAL ANALYSIS; ELEMENTS; EVALUATION; INFORMATION; METALS