Influence of amplifier time constant on proportional counter calibration

Creators

Description

This note describes the observation of the effects of a particular electronic parameter on the calibration of proportional counters. The effect in question is well-known and in a sense obvious, but it is easily overlooked in some applications of proportional counters to microdosimetry. Therefore, it is thought worthwhile to recount an example where this problem did occur, with a view toward helping others avoid such difficulties in the future. The effect will be of greatest importance for large gas pressures and/or small electrode potentials (low gas gain), and for radiation modalities which exhibit a large dynamic range of energy deposition

Additional details

Publishing Information

Imprint Title
Radiation physics, biophysics and radiation biology. Progress report, October 1, 1978-September 30, 1979
Journal Page Range
p. 72-75.
Report number
COO--4733-2

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
12589368
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
AMPLIFIERS; CALIBRATION; MICRODOSIMETRY; NEUTRON DOSIMETRY; PROPORTIONAL COUNTERS; TIME DEPENDENCE
Descriptors DEC
DOSIMETRY; MEASURING INSTRUMENTS; RADIATION DETECTORS