The build-up of backscatter in the energy range 1mm Al to 8mm Al HVT

  • 1. London Hospital (UK)

Description

Little information is available on the influence of the thickness of underlying material on the magnitude of the backscatter factor and investigations were therefore under-taken. Radiation beam qualities between 1mm Al to 8mm Al HVT and source-to-surface distances applicable to contemporary superficial radiotherapy were used. Measurements of the backscatter were made in a polystyrene phantom using a purpose-designed parallel-plate ionisation chamber built into the phantom surface. The effect of the thickness of underlying material on the build-up was investigated as a function of field size, beam quality, and SSD. The backscatter build-up curves were found to rise more slowly with increasing field size, increasing HVT and longer SSD; and the experimental data were fitted by an exponential function. A set of equations is proposed correlating the parameters influencing the backscatter behaviour and allowing the fractional backscatter factor to be calculated for a given field size, HVT, SSD and thickness of underlying material. (author)