Published September 1, 1996 | Version v1
Journal article

An approach to dose measurement for total body irradiation

Description

Purpose: An approach is proposed to allow the interpretation of diode measurements on patients receiving total body irradiation (TBI) in opposed lateral fields as midplane dose measurements. Methods and Materials: This technique consists of making measurements on both entrance and exit sides of any ray path along which the midplane dose is desired. The diode on the entrance side is calibrated so that its output is indicative of the dose delivered at the depth of maximum dose along the ray. The diode on the exit side is calibrated so that its output represents the dose that would have been delivered to that depth had the medium (the patient) been semi-infinite. Because these two measurements lie along the same ray, they are related by a simple depth-dose equation from which the effective absorption coefficient (μeff) can be determined. The dose at the midplane is calculated using this μeff in a similar equation. Results: The validity of this approach was checked with measurements at isocenter and at the TBI distance using a polystyrene phantom. In both cases, the calculated midplane dose was within reasonable experimental error (range ∼±2% on average) of the measured dose. Measurements on eight patients showed the excellent reproducibility of entrance surface measurements (standard deviation [SD] ∼±1%), compared to that of midplane measurements (SD ∼±4%) and exit surface measurements (SD ∼≤8%). Midplane doses determined from these equations confirmed that dose inhomogeneity using this opposed lateral technique is on the order of ±10% with highest doses delivered to the calves and lowest to the lungs. Conclusion: Recent measurements made on TBI patients have shown that a commercial eight-diode system can be used successfully to measure the dose delivered to the midplane within an accuracy of ∼±4%. The reproducibility of entrance surface measurements (∼±1%) shows that such measurements provide the accuracy required for quality assurance purposes

Additional details

Identifiers

PII
S0360301696002684;

Publishing Information

Journal Title
International Journal of Radiation Oncology, Biology and Physics
Journal Volume
36
Journal Issue
2
Journal Page Range
p. 463-468
ISSN
0360-3016
CODEN
IOBPD3

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35009122
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
DEPTH DOSE DISTRIBUTIONS; QUALITY ASSURANCE; RADIATION DOSES; RADIOTHERAPY; SEMICONDUCTOR DIODES; WHOLE-BODY IRRADIATION
Descriptors DEC
DOSES; EXTERNAL IRRADIATION; IRRADIATION; MEDICINE; NUCLEAR MEDICINE; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; SEMICONDUCTOR DEVICES; SPATIAL DOSE DISTRIBUTIONS; THERAPY

Optional Information

Copyright
Copyright (c) 1996 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.