Comparing five different iterative reconstruction algorithms for computed tomography in an ROC study
Creators
- 1. University of Oslo, lnstitute of Physics, Oslo (Norway)
- 2. Rikshospitalet, The Intervention Centre, Postboks 4950, Oslo (Norway)
- 3. Lund University, Skaane University Hospital, Department of Medical Radiation Physics, Malmoe (Sweden)
- 4. Rikshospitalet, Department of Radiology and Nuclear Medicine, Postboks 4950, Oslo (Norway)
- 5. University of Oslo, lnstitute of Clinical Medicine, Oslo (Norway)
Description
The purpose of this study was to evaluate lesion conspicuity achieved with five different iterative reconstruction techniques from four CT vendors at three different dose levels. Comparisons were made of iterative algorithm and filtered back projection (FBP) among and within systems. An anthropomorphic liver phantom was examined with four CT systems, each from a different vendor. CTDIvol levels of 5 mGy, 10 mGy and 15 mGy were chosen. Images were reconstructed with FBP and the iterative algorithm on the system. Images were interpreted independently by four observers, and the areas under the ROC curve (AUCs) were calculated. Noise and contrast-to-noise ratios (CNR) were measured. One iterative algorithm increased AUC (0.79, 0.95, and 0.97) compared to FBP (0.70, 0.86, and 0.93) at all dose levels (p < 0.001 and p = 0.047). Another algorithm increased AUC from 0.78 with FBP to 0.84 (p = 0.007) at 5 mGy. Differences at 10 and 15 mGy were not significant (p-values: 0.084-0.883). Three algorithms showed no difference in AUC compared to FBP (p-values: 0.008-1.000). All of the algorithms decreased noise (10-71 %) and improved CNR. Only two algorithms improved lesion detection, even though noise reduction was shown with all algorithms. (orig.)
Availability note (English)
Available from: http://dx.doi.org/10.1007/s00330-014-3333-4Additional details
Identifiers
Publishing Information
- Journal Title
- European Radiology
- Journal Volume
- 24
- Journal Issue
- 12
- Journal Page Range
- p. 2989-3002
- ISSN
- 0938-7994
- CODEN
- EURAE3
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- Germany
- INIS RN
- 46006964
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ABDOMEN; ABSORBED RADIATION DOSES; ALGORITHMS; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; IMAGE PROCESSING; ITERATIVE METHODS; LIVER; MILLI GY RANGE 01-10; MILLI GY RANGE 10-100; PHANTOMS; SIGNAL-TO-NOISE RATIO; STATISTICS
- Descriptors DEC
- ABSORBED DOSE RANGE; BODY; CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; DIGESTIVE SYSTEM; DIMENSIONLESS NUMBERS; DOSES; EVALUATION; GLANDS; MATHEMATICAL LOGIC; MATHEMATICS; MILLI GY RANGE; MOCKUP; ORGANS; PROCESSING; RADIATION DOSE RANGES; RADIATION DOSES; STRUCTURAL MODELS; TOMOGRAPHY