Published May 1976 | Version v1
Journal article

Electron scattering and collimation system for a 12-MeV linear accelerator

  • 1. Joint Center for Radiation Therapy and Department of Radiation Therapy, Harvard Medical School, Boston, Massachusetts 02115

Description

The original scattering and collimation system for the Siemens Mevatron XII linear accelerator used a lead scattering foil and box-type plastic collimators. This arrangement achieves excellent field flatness by repeated electron scattering. The electrons reaching the patient are widely distributed in energy and direction. This has detrimental effects on the depth-dose curves: slower falloff and increased surface dose. An alternative system was developed for this accelerator, designed to minimize electron scatter and improve the safety of patient setup. Primary-electron scatter occurs in the bending-magnet exit window. Field uniformity is accomplished with a flattener of thin aluminum discs of different diameters, piled concentrically. An adjustable electron collimator 25 cm from the patient limits beam size, and a final electron collimator, either a cutout from lead sheet or a a custom-made collimator of Lipowitz's metal, in contact with the patient, define the area to be treated. This design results in lower surface dose, sharper dose falloff, bremsstrahlung contamination < or =1%, and a field flatness expressed by a homogeneity index >0.8 for large fields. Since there is no mechanical connection between the machine and the final collimator, the safety aspects of the system are considerably improved

Additional details

Identifiers

Publishing Information

Journal Title
Medical Physics
Journal Volume
3
Journal Issue
3
Series
Med. Phys.
Journal Page Range
153-158
ISSN
0094-2405

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
7268894
Subject category
S43: PARTICLE ACCELERATORS;
Descriptors DEI
COLLIMATORS; DEPTH DOSE DISTRIBUTIONS; ELECTRON BEAMS; LINEAR ACCELERATORS; RADIOTHERAPY; SAFETY; SCATTERING
Descriptors DEC
ACCELERATORS; BEAMS; CYCLIC ACCELERATORS; LEPTON BEAMS; MEDICINE; PARTICLE BEAMS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DOSE DISTRIBUTIONS; THERAPY

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent