Published October 20, 2008 | Version v1
Journal article

Aqueous blackbody calibration source for millimeter-wave/terahertz metrology

Description

This paper describes a calibrated broadband emitter for the millimeter-wave through terahertz frequency regime, called the aqueous blackbody calibration source. Due to its extremely high absorption, liquid water is chosen as the emitter on the basis of reciprocity. The water is constrained to a specific shape (an optical trap geometry) in an expanded polystyrene (EPS) container and maintained at a selected, uniform temperature. Uncertainty in the selected radiometric temperature due to the undesirable reflectance present at a water interface is minimized by the trap geometry, ensuring that radiation incident on the entrance aperture encounters a pair of s and a pair of p reflections at 45 deg. . For water reflectance Rw of 40% at 45 deg. in W-band, this implies a theoretical effective aperture emissivity of (1-R2wsR2wp)>98.8%. From W-band to 450 GHz, the maximum radiometric temperature uncertainty is ±0.40 K, independent of water temperature. Uncertainty from 450 GHz to 1 THz is increased due to EPS scattering and absorption, resulting in a maximum uncertainty of -3 K at 1 THz

Additional details

Identifiers

Publishing Information

Journal Title
Applied Optics
Journal Volume
47
Journal Issue
30
Journal Page Range
p. 5604-5615
ISSN
0003-6935
CODEN
APOPAI

Optional Information

Notes
(c) 2008 Optical Society of America