Effect of scatter and an antiscatter grid on the performance of a slot-scanning digital mammography system
Creators
- 1. Division of Medical Physics, CancerCare Manitoba, 675 McDermot Ave Winnipeg, Manitoba, R3E 0V9 (Canada)
- 2. Imaging Research, Sunnybrook and Women's College Health Sciences Centre, 2075 Bayview Ave, Room SG12, Toronto, Ontario, M4N 3M5 (Canada)
Description
The use of a grid increases perceptibility of low contrast objects in mammography. Slot-scan mammography provides a more dose efficient reduction of the scattered radiation reaching the detector than obtained with an antiscatter grid in screen-film or flat-panel digital mammography. In this paper, the potential of using a grid in a slot-scan system to provide a further reduction of scattered radiation is investigated. The components of the digital signal: primary radiation, off-focus radiation, scattered radiation, and optical fluorescence glare in a CsI(Tl) detector were quantified. Based on these measurements, the primary and scatter transmission factors (Tp,Ts), scatter-to-primary ratio (SPR), signal-difference-to-noise ratio (SDNR), and the SDNR improvement factor (KSDNR) were obtained. Our results showed that the SPR ranged from 0.05 to 0.19 for breast thicknesses between 2 and 8 cm, respectively. The values of KSDNR ranged from 0.85 to 0.94. Because the slot-scanning system has an inherently low SPR, the increase in dose required when the grid is used outweighs the benefit of the small increase in SDNR. It is possible that greater benefit could be achieved by using a grid with a higher Tp, such as obtained using air-core technology
Additional details
Identifiers
- DOI
- 10.1118/1.2184445;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 33
- Journal Issue
- 4
- Journal Page Range
- p. 1108-1115
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37078682
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; FLUORESCENCE; IMAGE SCANNERS; MAMMARY GLANDS; PERFORMANCE; RADIATION DOSES; X-RAY DIFFRACTION
- Descriptors DEC
- BODY; COHERENT SCATTERING; DIAGNOSTIC TECHNIQUES; DIFFRACTION; DOSES; EMISSION; GLANDS; LUMINESCENCE; MEDICINE; NUCLEAR MEDICINE; ORGANS; PHOTON EMISSION; RADIOLOGY; SCATTERING
Optional Information
- Notes
- (c) 2006 American Association of Physicists in Medicine