Published November 26, 2007
| Version v1
Journal article
Simultaneous detection of translational and angular displacements of micromachined elements
- 1. Department of Mechanical Engineering, National Taiwan University, Taipei, Taiwan (China)
- 2. Institute of Physics, Academia Sinica, Nankang, Taipei, Taiwan (China)
Description
An astigmatic detection system is constructed with a modified digital-versatile-disk optical head. This system, with a detecting spot of ∼1 μm, can simultaneously measure the vertical displacements and two-dimensional angular tilts of micromachined elements. It can detect thermal vibrations of microfabricated cantilevers with noise levels of 1.3 pm Hz-1/2 for the linear displacement and of 3.2 nrad Hz-1/2 for angular displacements over a frequency range from 1 to 800 kHz. The detecting frequency can even reach beyond 100 MHz if high-speed electronic devices are adopted. Further optimization of the system will broaden its applications in diverse technological fields
Additional details
Identifiers
- DOI
- 10.1063/1.2817750;
Publishing Information
- Journal Title
- Applied Physics Letters
- Journal Volume
- 91
- Journal Issue
- 22
- Journal Page Range
- p. 221908-221908.3
- ISSN
- 0003-6951
- CODEN
- APPLAB
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 39038401
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ELECTRONIC EQUIPMENT; KHZ RANGE; MACHINING; MHZ RANGE 100-1000; MICROSTRUCTURE; OPTIMIZATION; TWO-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- EQUIPMENT; FREQUENCY RANGE; MHZ RANGE
Optional Information
- Notes
- (c) 2007 American Institute of Physics