Published May 1987
| Version v1
Journal article
Minimization of the blank values in the neutron activation analysis of biological samples considering the whole procedure
Creators
- 1. Technische Univ. Muenchen (Germany, F.R.). Inst. fuer Radiochemie
Description
During the determination of trace element contents of animal tissue by neutron activation analysis in the course of structure-activity relationship studies on platinum containing cancer drugs and wound healing the authors tried to minimize the blank values that are caused by different sources of contamination during surgery, sampling and the activation analysis procedure. The following topics were investigated: abrasions from scalpels made of stainless steel, titanium or quartz, the type of surgery, the surface contaminations of the quartz ampoules, etc. The measures to be taken to reduce tha blank values are described. (author) 19 refs.; 4 tables
Additional details
Publishing Information
- Journal Title
- J. Radioanal. Nucl. Chem.
- Journal Volume
- 112
- Journal Issue
- 1
- Series
- J. Radioanal. Nucl. Chem. ;Articles.
- Journal Page Range
- 161-168
- ISSN
- 0236-5731
- CODEN
- JRNCD
Conference
- Title
- 7. International conference on modern trends in activation analysis.
- Dates
- 23-27 Jun 1986.
- Place
- Copenhagen (Denmark).
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 19007557
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Conference, Numerical Data
- Descriptors DEI
- ABRASION; BIOLOGICAL MATERIALS; CONTAMINATION; EXPERIMENTAL DATA; MULTI-ELEMENT ANALYSIS; NEUTRON ACTIVATION ANALYSIS; SAMPLE PREPARATION; SURGICAL MATERIALS; TISSUES; TRACE AMOUNTS
- Descriptors DEC
- ACTIVATION ANALYSIS; BODY; CHEMICAL ANALYSIS; DATA; INFORMATION; MATERIALS; MEDICAL SUPPLIES; NONDESTRUCTIVE ANALYSIS; NUMERICAL DATA