Longitudinal and transverse digital image reconstruction with a tomographic scanner
- 1. Vanderbilt Univ., Nashville, TN (USA). Medical Center
Description
A Siemens Gammasonics PHO/CON-192 Multiplane Imager is interfaced to a digital computer for the purpose of performing tomographic reconstructions from the data collected during a single scan. Data from the two moving gamma cameras as well as camera position information are sent to the computer by an interface designed in the authors' laboratory. Backprojection reconstruction is implemented by the computer. Longitudinal images in whole-body format as well as smaller formats are reconstructed for up to six planes simultaneously from the list mode data. Transverse reconstructions are demonstrated for 201T1 myocardial scans. Post-reconstruction deconvolution processing to remove the blur artifact (characteristic of focal plane tomography) is applied to a multiplane phantom. Digital data acquisition of data and reconstruction of images are practical, and can extend the usefulness of the machine when compared with the film output (author)
Files
12642456.pdf
Files
(460.7 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:d01fb4e1768dcd09bebb29a85ba65132
|
460.7 kB | Preview Download |
Additional details
Publishing Information
- Publisher
- IAEA.
- Imprint Place
- Vienna
- ISBN
- 92-0-010081-3
- Imprint Title
- Medical radionuclide imaging 1980
- Imprint Pagination
- 635 p.
- Series
- Proceedings series.
- Journal Page Range
- v. 1 p. 325-332.
Conference
- Title
- International symposium on medical radionuclide imaging.
- Dates
- 1 - 5 Sep 1980.
- Place
- Heidelberg, Germany, F.R.
INIS
- Country of Publication
- International Atomic Energy Agency (IAEA)
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 12642456
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DATA ACQUISITION; GAMMA CAMERAS; IMAGE PROCESSING; IMAGES; MYOCARDIUM; PHANTOMS; RADIOISOTOPE SCANNERS; SCINTISCANNING; THALLIUM 201; TOMOGRAPHY
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BODY; CAMERAS; CARDIOVASCULAR SYSTEM; COUNTING TECHNIQUES; DAYS LIVING RADIOISOTOPES; DIAGNOSTIC TECHNIQUES; ELECTRON CAPTURE RADIOISOTOPES; HEART; HEAVY NUCLEI; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MOCKUP; MUSCLES; NUCLEI; ODD-EVEN NUCLEI; ORGANS; RADIOISOTOPE SCANNING; RADIOISOTOPES; SECONDS LIVING RADIOISOTOPES; STRUCTURAL MODELS; THALLIUM ISOTOPES
Optional Information
- Secondary number(s)
- IAEA-SM--247/76.