Published May 1995 | Version v1
Journal article

Progress at the Groningen AMS facility

  • 1. High Voltage Eng. Europa B.V., Amersfoort (Netherlands)
  • 2. Centre for Isotope Research, University of Groningen (Netherlands)

Description

A new generation accelerator mass spectrometer (AMS) is operational since May 1994 at the Centre for Isotope Research in Groningen, The Netherlands. The fully automated and high throughput AMS system, manufactured by High Voltage Engineering Europa (HVEE) is dedicated to radiocarbon analysis. The model 4130 14C AMS is able to analyse up to 3000 samples per year. The system is characterized by a simultaneous transport of all three isotopes. Since the system is fully operational, various tests have been done to obtain insight in the system performance. Background, including sample preparation is, at present, approximately 45 ka and is still improving, whereas machine background is negligible. Performance tests on identical samples and cross checking with conventional 14C decay counting and 13C spectrometry showed a precision on the 13δ and the 14C/12C ratio of better than 2 permille and 0.5 pMC respectively. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section B, Beam Interactions with Materials and Atoms
Journal Volume
99
Journal Issue
1-4
Journal Page Range
p. 532-536.
ISSN
0168-583X
CODEN
NIMBEU

Conference

Title
13. international conference on the application of accelerators in research and industry - topical conference of the Division of Nuclear Physics (DNP) of the American Physical Society (APS).
Dates
7-10 Nov 1994.
Place