Establishment of local diagnostic reference levels for paediatric abdominal-pelvis and chest-abdominal-pelvis computed tomography in Morocco: suggests the need for improved optimization efforts

  • 1. Sidi Mohamed Ben Abdellah University, Faculty of Sciences Dhar EL Mahraz, Fez (Morocco)
  • 2. Sidi Mohamed Ben Abdellah University, Higher School of Technology, Fez (Morocco)
  • 3. Hassan First University, Higher Institute of Health Sciences, Settat (Morocco)
  • 4. University Mohammed First, National School of Applied Sciences, Oujda (Morocco)

Description

The purpose of the current study was to derive the local diagnostic reference levels (LDRLs) for paediatric abdominal-pelvis (AP) and chest-abdominal-pelvis (CAP) computed tomography in Morocco. The data were gathered retrospectively from two hospitals for 6 months. The LDRLs were defined by volume CT dose index (CTDIvol), dose-length product (DLP) per sequence, DLP per procedure and size-specific dose estimates (ssDE). The ssDE assessment was based on the effective diameters of patients scanned. A total of 630 CT examinations were collected involving 324 AP and 306 CAP scans. The proposed LDRLs for AP, in terms of CTDIvol (mGy), were 6.9, 8.5, 8.5 and 8.5 for < 1, 1 to < 5, 5 to < 10 and 10 to < 15 y age groups, respectively. In terms of DLP (mGy.cm) per procedure, they were 436.3, 534.5, 687.9 and 961.7. In terms of ssDE (mGy), they were 16.73, 16.83, 17.5 and 15.8 for < 1, 1 to < 5, 5 to < 10 and 10 to < 15 y, respectively. The corresponding LDRLs for CAP, in terms of CTDIvol (mGy), were 7.3, 7.3, 7.3 and 10.35. In terms of DLP (mGy.cm) per procedure, they were 531, 622.5, 705 and 936. In terms of ssDE (mGy), they were 16.22, 15.05, 14.47 and 15.2, respectively, for the four age groups. The derived dose levels were mostly higher than those found in other studies, which demonstrates the need for dose optimization and paediatric protocol standardization as well as the timeliness of the intent to establish not only local DRLs but national ones in the near future. (authors)

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
199
Journal Issue
10
Journal Page Range
p. 1129-1141
ISSN
0144-8420

INIS

Country of Publication
United Kingdom
Country of Input or Organization
France
INIS RN
55005540
Subject category
S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
CHEST; COMPUTERIZED TOMOGRAPHY; OPTIMIZATION; PATIENTS; PELVIS; RADIATION DOSES; STANDARDIZATION; TIME DELAY
Descriptors DEC
BODY; DIAGNOSTIC TECHNIQUES; DOSES; TOMOGRAPHY

Optional Information

Notes
40 refs.