A time-of-flight flow sensor for the volume measurement of trace amount of interstitial fluid
Creators
- 1. State Key Laboratory of Precision Measuring Technology and Instruments, Tianjin University, Tianjin 300072 (China)
- 2. Department of Electrical Engineering and Computer Science, Case Western Reserve University, Cleveland, OH 44106 (United States)
Description
Transdermal extraction of interstitial fluid (ISF) offers an attractive method for noninvasive blood glucose monitoring. The existing macroscale systems are not suitable for ISF collection, mainly because of the minute volume of the transdermally extracted ISF which scatters on the skin surface. Human skin's low permeability to glucose and its varying permeability exemplify the crucial need to make precise ISF volume measurements in order to calculate blood glucose concentrations accurately. In this paper, we present a novel time-of-flight flow sensor consisting of four electrode pairs fabricated directly into the channel of a polydimethylsiloxane (PDMS) microfluidic device designed to accurately measure the volume of transdermally extracted ISF. As fluid traverses the channel, it bridges each electrode pair and changes its resistance. By measuring the time difference in resistance change between each electrode pair, a precise fluid volume can be measured. In order to verify the suitability of the sensor for biological applications, experiments were conducted using a normal saline solution which is similar to ISF. The stability of the sensor was tested using a fixed volume, and the coefficient of variation for 20 tests was determined to be 0.0041. The consistency of the sensor for varied volume measurements was shown by the high correlation coefficient (R2 = 0.9992) between the tested volume and the volume measured by a commercial micro syringe. The excellent functionality of the flow sensor can be extended toward the measurement of conductive chemical and biochemical buffers and reagents. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0960-1317/22/5/055009Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Micromechanics and Microengineering. Structures, Devices and Systems
- Journal Volume
- 22
- Journal Issue
- 5
- Journal Page Range
- [7 p.]
- ISSN
- 0960-1317
- CODEN
- JMMIEZ
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47014247
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY; S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Descriptors DEI
- ABUNDANCE; BLOOD; BUFFERS; CONCENTRATION RATIO; ELECTRODES; EXTRACTION; FLUIDS; GLUCOSE; HUMAN POPULATIONS; INTERSTITIALS; PERMEABILITY; REAGENTS; SENSORS; SKIN; TIME-OF-FLIGHT METHOD; TRACE AMOUNTS
- Descriptors DEC
- ALDEHYDES; BIOLOGICAL MATERIALS; BODY; BODY FLUIDS; CARBOHYDRATES; CRYSTAL DEFECTS; CRYSTAL STRUCTURE; DIMENSIONLESS NUMBERS; HEXOSES; MATERIALS; MONOSACCHARIDES; ORGANIC COMPOUNDS; ORGANS; PHYSICAL PROPERTIES; POINT DEFECTS; POPULATIONS; SACCHARIDES; SEPARATION PROCESSES