Published January 2020 | Version v1
Journal article

Determination of ionic 68Ga impurity in radiopharmaceuticals. Major revision of radio-HPLC methods

  • 1. Russian Academy of Sciences, Moscow (Russian Federation). Frumkin Institute of Physical Chemistry and Electrochemistry
  • 2. Russian State Research Center - Burnasyan Federal Medical Biophysical Center of Federal Medical Biological Agency, Moscow (Russian Federation)
  • 3. Lomonosov Moscow State University, Moscow (Russian Federation)

Description

Determination of purity of 68Ga-radiopharmaceuticals is an extremely important part of quality control in routine clinical practice as well as during R&D of 68Ga-radiopharmaceuticals. HPLC results do not always match TLC results. This uncertainty most likely comes from nonspecific sorption of ionic 68Ga on C18 phase. The aim of this study was to develop reliable HPLC analysis procedure. It was shown that simple replacement of trifluoroacetic acid in the eluent with citric acid results into change of the results obtained using HPLC analysis. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
323
Journal Issue
1
Journal Page Range
p. 189-195
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
22 refs.