Published August 2020 | Version v1
Journal article

Comparison of gamma index passing rate in several treatment planning system algorithms

  • 1. Department of Radiotherapy, MRCCC Siloam Hospitals, Jl. Garnisun No.5, Semanggi, Jakarta 12930 (Indonesia)
  • 2. Department of Physics, Faculty of Mathematics and Natural Sciences, University of Indonesia, Jl. Margonda Raya, Depok 16424 (Indonesia)

Description

The verification of dose calculation algorithm in a new Treatment Planning System (TPS) can be evaluated by comparing the passing rate of the gamma index analysis of the evaluated algorithm and the clinically implemented algorithms. In the present investigation, the gamma index passing rates was investigated as the reference data in the verification of the new three-dimensional TPS. The algorithms which are used in this study are Pencil Beam Convolution (PBC) version 11.0.31 and Anisotropic Analytical Algorithm (AAA) version 11.0.31 in Eclipse v.11 TPS, and Fast Convolution (FC), Adaptive Convolution (AC), and Collapsed-Cone Convolution (CCC) in Pinnacle3 v.7.6c TPS. The 6 MV X-ray beam configurations were varied in the depth of measurement points, field sizes, source-to-surface distances, and wedge angles. The dose measurement was done using MatriXX Evolution and PTW 2D-array seven29. Then, OmniPro ImRT and Verisoft 3.1 software were chosen to analyze the gamma index from varied gamma criteria (3 %/3 mm, 2 %/3 mm, 3 %/2 mm, and 2 %/2 mm). Overall, the passing rate of AAA is the highest rate obtained of all algorithms. For gamma criterion of 2 %/2 mm, the passing rate of AAA was 93.18 % ± 7.21 %, the passing rate of PBC was 89.76 % ± 7.21 %, and the passing rate of convolution algorithms was 76.84 % ± 11.10 %. (author)

Additional details

Publishing Information

Journal Title
Atom Indonesia
Journal Volume
46
Journal Issue
2
Journal Page Range
p. 77-84
ISSN
0126-1568
CODEN
ATINDD

INIS

Country of Publication
Indonesia
Country of Input or Organization
Indonesia
INIS RN
52031474
Subject category
S97: MATHEMATICAL METHODS AND COMPUTING;
Resource subtype / Literary indicator
Numerical Data
Descriptors DEI
ACCURACY; ALGORITHMS; BEAMS; COMPUTER CALCULATIONS; DOSE EQUIVALENTS; EVALUATED DATA; GAMMA FUNCTION; INDEXES; VERIFICATION
Descriptors DEC
DATA; DOCUMENT TYPES; FUNCTIONS; INFORMATION; MATHEMATICAL LOGIC; NUMERICAL DATA

Optional Information

Notes
20 refs.; 3 tabs.; 5 figs