Published November 2018 | Version v1
Report

Development of measurement method of radiostrontium using plastic scintillator bottle

  • 1. Hitachi, Ltd., Measuring Systems Design Dept., Tokyo (Japan)
  • 2. Jikei University School of Medicine, Radioisotope Research Facility, Tokyo (Japan)
  • 3. Nagoya University, Radioisotope Research Center Medical Division, Nagoya, Aichi (Japan)

Description

Rapid and convenient analysis of radiostrontium is desired after the nuclear accident at Fukushima Daiichi Nuclear Power Plant. The conventional method to separate radiostrontium in seawater is time-consuming complicated procedures with deleterious substances. In this study, a new method to measure radiostrontium in seawater was developed using plastic scintillator via liquid scintillation system coupled with a convenient chemical separation technique. The counting efficiency was 70%, the background was 0.14 cps. When the counting time is 60 minutes and 70% of chemical yield, the minimum detectable concentration is estimated as 0.02 Bq L-1 started with 1 L water sample. This method is proved to be effective for rapid screening of radioactive strontium in seawater. (author)

Part of:
Proceedings of the 19th workshop on environmental radioactivity

Additional details

Additional titles

Original title (Japanese)
プラスチックシンチレータボトルを用いた放射性ストロンチウムの測定法の開発

Publishing Information

Imprint Title
Proceedings of the 19th workshop on environmental radioactivity
Imprint Pagination
340 p.
Journal Page Range
p. 85-90
Report number
KEK-PROC--2018-7

Conference

Title
19. workshop on environmental radioactivity
Dates
13-15 Mar 2018
Place
Tsukuba, Ibaraki (Japan)

Optional Information

Notes
3 refs., 9 figs., 2 tabs. Imprint:#7B2C#19#56DE##300C##74B0##5883##653E##5C04##80FD##300D##7814##7A76##4F1A#