Published 1990
| Version v1
Journal article
Comparison of four commercially available 81Rb-81mKr generators: quality testing using a breathing lung phantom
- 1. University Hospital, Leiden (Netherlands). Div. of Nuclear Medicine
Description
Four commercially available 81Rb-81mKr generators, including their elution systems, were compared. Considerable differences in useful output between the various generators were found. When a standardized elution system was used, the differences became much less. Differences in the tubing and inaccuracy of the flow meters were found to be the major causes of the original differences. The user should be aware of a loss of output that may occur when the generator is eluted with dry air. (author)
Additional details
Publishing Information
- Journal Title
- Nuclear Medicine and Biology
- Journal Volume
- 17
- Journal Issue
- 3
- Series
- Nucl. Med. Biol.
- Journal Page Range
- 269-272
- ISSN
- 0883-2897
- CODEN
- NMBIE
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 21075606
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COMPARATIVE EVALUATIONS; HUMIDITY; KRYPTON 81; LUNGS; PERFORMANCE; PHANTOMS; RADIOISOTOPE GENERATORS; RUBIDIUM 81
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-PLUS DECAY RADIOISOTOPES; BODY; ELECTRON CAPTURE RADIOISOTOPES; EVALUATION; EVEN-ODD NUCLEI; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTO; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; KRYPTON ISOTOPES; MINUTES LIVING RADIOISOTOPES; MOCKUP; NUCLEI; ODD-EVEN NUCLEI; ORGANS; RADIOISOTOPES; RESPIRATORY SYSTEM; RUBIDIUM ISOTOPES; SECONDS LIVING RADIOISOTOPES; STRUCTURAL MODELS; YEARS LIVING RADIOISOTOPES