SU-E-T-105: Development of 3D Dose Verification System for Volumetric Modulated Arc Therapy Using Improved Polyacrylamide-Based Gel Dosimeter
- 1. Hiroshima Heiwa Clinic, Hiroshima (Japan)
- 2. Hiroshima International University, Hiroshima (Japan)
- 3. R-TECH.INC, Toukyo (Japan)
Description
Purpose: The aim of this dosimetric study was to develop 3D dose verification system for volumetric modulated arc therapy (VMAT) using polyacrylamide-based gel (PAGAT) dosimeter improved the sensitivity by magnesium chloride (MgCl2). Methods: PAGAT gel containing MgCl2 as a sensitizer was prepared in this study. Methacrylic-acid-based gel (MAGAT) was also prepared to compare the dosimetric characteristics with PAGAT gel. The cylindrical glass vials (4 cm diameter, 12 cm length) filled with each polymer gel were irradiated with 6 MV photon beam using Novalis Tx linear accelerator (Varian/BrainLAB). The irradiated polymer gel dosimeters were scanned with Signa 1.5 T MRI system (GE), and dose calibration curves were obtained using T2 relaxation rate (R2 = 1/T2). Dose rate (100-600 MU min−1) and fractionation (1-8 fractions) were varied. In addition, a cubic acrylic phantom (10 × 10 × 10 cm3) filled with improved PAGAT gel inserted into the IMRT phantom (IBA) was irradiated with VMAT (RapidArc). C-shape structure was used for the VMAT planning by the Varian Eclipse treatment planning system (TPS). The dose comparison of TPS and measurements with the polymer gel dosimeter was accomplished by the gamma index analysis, overlaying the dose profiles for a set of data on selected planes using in-house developed software. Results: Dose rate and fractionation dependence of improved PAGAT gel were smaller than MAGAT gel. A high similarity was found by overlaying the dose profiles measured with improved PAGAT gel dosimeter and the TPS dose, and the mean pass rate of the gamma index analysis using 3%/3 mm criteria was achieved 90% on orthogonal planes for VMAT using improved PAGAT gel dosimeter. Conclusion: In-house developed 3D dose verification system using improved polyacrylamide-based gel dosimeter had a potential as an effective tool for VMAT QA
Additional details
Identifiers
- DOI
- 10.1118/1.4888435;
Publishing Information
- Journal Title
- Medical Physics
- Journal Volume
- 41
- Journal Issue
- 6
- Journal Page Range
- p. 246
- ISSN
- 0094-2405
- CODEN
- MPHYA6
INIS
- Country of Publication
- United States
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 46046177
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- AMINO ACIDS; DOSE RATES; DOSEMETERS; GELS; LINEAR ACCELERATORS; MAGNESIUM CHLORIDES; METHACRYLIC ACID; NMR IMAGING; PHANTOMS; PHOTON BEAMS; RADIATION DOSES; RADIOTHERAPY; VERIFICATION
- Descriptors DEC
- ACCELERATORS; ALKALINE EARTH METAL COMPOUNDS; BEAMS; CARBOXYLIC ACIDS; CHLORIDES; CHLORINE COMPOUNDS; COLLOIDS; DIAGNOSTIC TECHNIQUES; DISPERSIONS; DOSES; HALIDES; HALOGEN COMPOUNDS; MAGNESIUM COMPOUNDS; MAGNESIUM HALIDES; MEASURING INSTRUMENTS; MEDICINE; MOCKUP; MONOCARBOXYLIC ACIDS; NUCLEAR MEDICINE; ORGANIC ACIDS; ORGANIC COMPOUNDS; RADIOLOGY; STRUCTURAL MODELS; THERAPY
Optional Information
- Notes
- (c) 2014 American Association of Physicists in Medicine