Published March 1985
| Version v1
Journal article
The energizing of a NMR superconducting coil by a superconducting rectifier
Creators
- 1. Holec Innovation and Technology, Hengelo
Description
NMR magnets require a good homogeneity within a certain volume and an excellent field stability. The homogeneity can be met using a superconducting shim coil system. The field stability requires a constant current, although in many cases the current decay time constant is too low, due to imperfections in the superconducting wire and joints. This can be overcome using a rectifier. The rectifier can also be used to load the coil. The combination and interaction of the superconducting NMR coil (2.0 Tesla and 0.35 m cold bore) and the rectifier (20 W / 1 kA) is tested. The safety of the system is discussed. The shim coil system can compensate the strayfield of the rectifier. The field decay compensation will be discussed
Additional details
Publishing Information
- Journal Title
- IEEE Trans. Magn.
- Journal Volume
- MAG 21
- Journal Issue
- 2
- Series
- IEEE Trans. Magn.
- Journal Page Range
- 440-443
- ISSN
- 0018-9464
- CODEN
- IEMGA
Conference
- Title
- Applied superconductivity conference.
- Dates
- 9-13 Sep 1984.
- Place
- San Diego, CA (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 17030229
- Subject category
- S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- DEFECTS; EXCITATION; MAGNETIC FIELDS; NUCLEAR MAGNETIC RESONANCE; OCCUPATIONAL SAFETY; PERFORMANCE TESTING; RECTIFIERS; STABILITY; SUPERCONDUCTING MAGNETS; SUPERCONDUCTING WIRES
- Descriptors DEC
- ELECTRICAL EQUIPMENT; ELECTROMAGNETS; ENERGY-LEVEL TRANSITIONS; EQUIPMENT; MAGNETIC RESONANCE; MAGNETS; RESONANCE; SAFETY; SUPERCONDUCTING DEVICES; TESTING; WIRES
Optional Information
- Secondary number(s)
- CONF-840937--.