The thickness of Cu filter to reduce 1/2 of the patient dose

  • 1. College of Health Sciences, Korea Univ., Seoul (Korea, Republic of)
  • 2. Gachun Gil College, Incheon (Korea, Republic of)

Description

Medical X-ray examination are increased double for the last 6 ∼ 8 years. Therefore a patient exposure dose should be decrease half every 7 years. We made an experiment on copper filter thickness to decrease a patient exposure dose up to half and compared to the image quality by MTF. The results as follow a thin region like extremities needs a thicker Cu filter as compared a thick region. 1/2 reduction filter must be thicker when kVp increase. Exposure factor should be increas when using 1/2 reduction filter : extremity is 4.0 times, chest 2.9 times, skull 1.62 times, and abdomen 1.58 times. The MTF of using 1/2 reduction filter is lower than without filter. But no difference of visual image. 1/2 reduction filter compared with double speed screen showed almost same image quality

Additional details

Publishing Information

Journal Title
Journal of the Korean Society of Radiological Technology
Journal Volume
24
Journal Issue
1
Series
8 refs, 9 figs, 4 tabs
Journal Page Range
p. 17-22
ISSN
1226-2854

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
32067181
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COPPER; FILTERS; IMAGES; MEDICAL EXAMINATIONS; PATIENTS; RADIATION DOSES; X-RAY RADIOGRAPHY
Descriptors DEC
DOSES; ELEMENTS; INDUSTRIAL RADIOGRAPHY; MATERIALS TESTING; MEDICAL SURVEILLANCE; METALS; NONDESTRUCTIVE TESTING; TESTING; TRANSITION ELEMENTS