Description

In this work a comparison of two passive decoupling techniques for ultra-high field magnetic resonance imaging (MRI) is presented. An electromagnetic band-gap (EBG) structure and a single passive dipole are compared in terms of their capabilities to decouple two active dipole body array elements at 7 Tesla at an extremely short inter-element distance. Decoupling band as well as distribution of circularly polarized radiofrequency magnetic field inside a homogeneous phantom provided with these two techniques are compared. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/1092/1/012049

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
1092
Journal Issue
1
Journal Page Range
[4 p.]
ISSN
1742-6596

Conference

Title
International Conference on Metamaterials and Nanophotonic
Acronym
METANANO 2018
Dates
17-21 Sep 2018
Place
Sochi (Russian Federation)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
53023354
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
Resource subtype / Literary indicator
Conference
Descriptors DEI
DECOUPLING; MAGNETIC FIELDS; MAGNETIC RESONANCE; NMR IMAGING; PHANTOMS; RADIOWAVE RADIATION
Descriptors DEC
DIAGNOSTIC TECHNIQUES; ELECTROMAGNETIC RADIATION; MOCKUP; RADIATIONS; RESONANCE; STRUCTURAL MODELS