Published 1979 | Version v1
Report Open

The absolute calibration of semiconductor detectors in the Neutrino beam of CERN

Description

The report describes a method for the calibration of semiconductor detectors. A nuclear emulsion is exposed to charged particles (muons) immediately in front of the detector. The muons also scatter delta electrons which give traces in the emulsion. The traces can be counted under a microscope. The separation of the muons and delta electrons takes place by angular distribution. The muons are counted per area unit. The flow is related to the signal of the detector and an absolute counting is achieved. (G.B)

Availability note (English)

MF available from INIS under the Report Number.

Files

12585997.pdf

Files (2.2 MB)

Name Size Download all
md5:7eab0295bf9713fac73b8c4bdde9f8ae
2.2 MB Preview Download

Additional details

Additional titles

Original title (Swedish)
Absolutkalibrering av halvledardetektorer i neutronstraalen paa CERN

Publishing Information

Imprint Pagination
66 p.
Report number