Published November 2018 | Version v1
Journal article

Measurement of internal radiation dose distribution in CT examinations using polyethylene terephthalate resin

  • 1. Department of Radiological Sciences, Graduate School of Medicine, Nagoya University, 1-1-20 Daikominami, Higashi-ku, Nagoya 461-8673 (Japan)
  • 2. Division of Radiology, Aichi Medical University Hospital, Nagakute, Aichi 480-1195 (Japan)
  • 3. Department of Radiological Technology, School of Health Sciences, Nagoya University, 1-1-20 Daikominami, Higashi-ku, Nagoya 461-8673 (Japan)

Description

This study proposes a new dosimetry method for the estimation of the internal radiation dose distribution of a subject undergoing computed tomography (CT) examinations. In this novel method, dose distribution of a subject by CT scans was estimated based on radio-photo-luminance distribution with polyethylene terephthalate (PET) resin which was cut to the average head size of a Japanese 1-year-old child. The difference in dose distribution depending on the type of bowtie filter was visualized by imaging luminance distribution with the PET phantom using a charge-coupled device camera. Dose distribution images simulated from a water phantom of the same size as the PET phantom were compared with the luminance distribution images. The linear correlation was demonstrated between luminance of the PET phantom and the simulated water dose. In comparison with the simulated water doses and the converted water doses from luminance of the PET phantom, the relative differences were within 20%. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncy026

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
181
Journal Issue
4
Journal Page Range
p. 303-309
ISSN
0144-8420

Optional Information

Notes
17 refs.