Published November 2003 | Version v1
Journal article

From PET detectors to PET scanners

  • 1. Department of Medical Physics, Memorial Sloan Kettering Cancer Center, 1275 York Avenue, NY 10021, New York (United States)
  • 2. Physics Department, University of Wollongong, New South Wales (Australia)
  • 3. Department of Physics ''E. Fermi'', University of Pisa, Pisa (Italy)

Description

This review describes the properties of available and emerging radiation detector and read-out technologies and discusses how they may affect PET scanner performance. After a general introduction, there is a section in which the physical properties of several different detector scintillators are compared. This is followed by a discussion of recent advances in read-out electronics. Finally, the physical performance of the several commercial PET scanners is summarized. (orig.)

Availability note (English)

Available from: http://dx.doi.org/10.1007/s00259-003-1266-2

Additional details

Identifiers

Publishing Information

Journal Title
European Journal of Nuclear Medicine and Molecular Imaging
Journal Volume
30
Journal Issue
11
Journal Page Range
p. 1574-1597
ISSN
1619-7070