3D-printing non-coplanar template assisted 125I seed implantation for pelvic tumor: individual template design method
Creators
- 1. Department of Radiation Oncology, Peking University Third Hospital, Beijing (China)
- 2. Image Processing Center, Beihang University, Beijing (China)
Description
Objective: To compare the dosimetric data between preoperative plans and postoperative verification in computed tomography (CT)-guided and 3D-printing non-coplanar template-assisted 125I seed implantation for pelvic tumor, and to explore the feasibility and accuracy of the personalized template designmethod. Methods: A total of 51 patients registered from Dec 2015 to Dec 2016 who were applied with 3D-printing guided template assisted radioactive seed implantations in the hospital were included in this study. A prescribed dose of 110-160 Gy was adopted. 3D-printing templates were designed and produced for 51 cases. The dosimetric parameters: D90, minimum peripheral dose (mPD), V100, V150, V200, conformal index (CI), external index (EI), and homogeneity index (HI) were compared between pre-and post-plans. Results: 51 cases' templates were in place well during the operations. Compared with the preoperative planning, the postoperative D90, V100, V150, V200, CI, EI and HI differences had no statistical difference (P > 0.05); mPD is larger than before (t = -2.96, P < 0.05). Conclusions: The main dosimetric parameters of postoperative verification were consistent well with the preoperative planning and have good accuracy, which could meet the clinical requirements. (authors)
Additional details
Identifiers
Publishing Information
- Journal Title
- Chinese Journal of Radiological Medicine and Protection
- Journal Volume
- 37
- Journal Issue
- 7
- Journal Page Range
- p. 485-489
- ISSN
- 0254-5098
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 50061321
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S97: MATHEMATICAL METHODS AND COMPUTING;
- Descriptors DEI
- ACCURACY; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; DESIGN; IODINE 125; MAXIMUM PERMISSIBLE DOSE; NEOPLASMS; PATIENTS; RADIATION DOSES; THREE-DIMENSIONAL LATTICES
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; CRYSTAL LATTICES; CRYSTAL STRUCTURE; DAYS LIVING RADIOISOTOPES; DIAGNOSTIC TECHNIQUES; DISEASES; DOSES; ELECTRON CAPTURE RADIOISOTOPES; EVALUATION; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; IODINE ISOTOPES; ISOTOPES; NUCLEI; ODD-EVEN NUCLEI; RADIOISOTOPES; SAFETY STANDARDS; STANDARDS; TOMOGRAPHY
Optional Information
- Notes
- 3 figs., 1 tab., 12 refs.; http://dx.doi.org/10.3760/cma.j.issn.0254-5098.2017.07.002