Published November 15, 1975
| Version v1
Journal article
E-dE/dX scintillator spectrometer for protons up to 120 MeV
Creators
- 1. Genoa Univ. (Italy). Istituto di Scienze Fisiche
Description
An E/dE-dx scintillation counter telescope for protons between 30 and 120 MeV was realized with a measured energy resolution of 2.5% (fwhm) on an 80 cm2 surface. The very fast analysis performed on the Na(Tl) counter pulses greatly reduces pile-up effects. (Auth.)
Additional details
Identifiers
Publishing Information
- Journal Title
- Nuclear Instruments and Methods
- Journal Volume
- 129
- Journal Issue
- 2
- Series
- Nucl. Instrum. Methods.
- Journal Page Range
- 467-472
- ISSN
- 0029-554X
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 7241171
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALIBRATION; EFFICIENCY; ENERGY RESOLUTION; MEV RANGE 10-100; MEV RANGE 100-1000; PROTON DETECTION; PROTON SPECTRA; PROTON SPECTROMETERS; PULSE PILEUP; SCINTILLATION COUNTERS; TELESCOPE COUNTERS
- Descriptors DEC
- CHARGED PARTICLE DETECTION; ENERGY RANGE; MEASURING INSTRUMENTS; MEV RANGE; RADIATION DETECTION; RADIATION DETECTORS; RESOLUTION; SPECTRA; SPECTROMETERS