Dose reconstruction from PET images in carbon ion therapy: a deconvolution approach
Creators
- 1. Department for Medical Physics, Ludwig-Maximilians-Universität München, Am Coulombwall 1, 85748 Garching (Germany)
- 2. Imaging Physics Team, National Institute of Radiological Sciences (NIRS-QST), 4-9-1 Anagawa, Inage, Chiba 263-8555 (Japan)
- 3. Australian Nuclear Science and Technology Organisation, Lucas Heights NSW, 2234 (Australia)
- 4. Center for Medical Radiation Physics, University of Wollongong, Wollongong NSW, 2522 (Australia)
Description
Dose and range verification have become important tools to bring carbon ion therapy to a higher level of confidence in clinical applications. Positron emission tomography is among the most commonly used approaches for this purpose and relies on the creation of positron emitting nuclei in nuclear interactions of the primary ions with tissue. Predictions of these positron emitter distributions are usually obtained from time-consuming Monte Carlo simulations or measurements from previous treatment fractions, and their comparison to the current, measured image allows for treatment verification. Still, a direct comparison of planned and delivered dose would be highly desirable, since the dose is the quantity of interest in radiation therapy and its confirmation improves quality assurance in carbon ion therapy.
In this work, we present a deconvolution approach to predict dose distributions from PET images in carbon ion therapy. Under the assumption that the one-dimensional PET distribution is described by a convolution of the depth dose distribution and a filter kernel, an evolutionary algorithm is introduced to perform the reverse step and predict the depth dose distribution from a measured PET distribution. Filter kernels are obtained from either a library or are created for any given situation on-the-fly, using predictions of the -decay and depth dose distributions, and the very same evolutionary algorithm.
The applicability of this approach is demonstrated for monoenergetic and polyenergetic carbon ion irradiation of homogeneous and heterogeneous solid phantoms as well as a patient computed tomography image, using Monte Carlo simulated distributions and measured in-beam PET data. Carbon ion ranges are predicted within less than 0.5 mm and 1 mm deviation for simulated and measured distributions, respectively. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1361-6560/aaf676Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 64
- Journal Issue
- 2
- Journal Page Range
- [15 p.]
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 52003617
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CARBON IONS; DEPTH DOSE DISTRIBUTIONS; FORECASTING; GENETIC ALGORITHMS; ION BEAMS; MONTE CARLO METHOD; POSITRON COMPUTED TOMOGRAPHY; QUALITY ASSURANCE; RADIATION DOSES; RADIOTHERAPY
- Descriptors DEC
- ALGORITHMS; BEAMS; CALCULATION METHODS; CHARGED PARTICLES; COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DOSES; EMISSION COMPUTED TOMOGRAPHY; IONS; MANAGEMENT; MATHEMATICAL LOGIC; MEDICINE; NUCLEAR MEDICINE; QUALITY MANAGEMENT; RADIATION DOSE DISTRIBUTIONS; RADIOLOGY; SPATIAL DOSE DISTRIBUTIONS; THERAPY; TOMOGRAPHY