Published January 5, 2006 | Version v1
Journal article

Validation of an online dual-loop cleanup device with an electrospray ionization tandem mass spectrometry-based system for simultaneous quantitative analysis of urinary benzene exposure biomarkers trans, trans-muconic acid and S-phenylmercapturic acid

  • 1. Department of Environmental and Occupational Health, College of Medicine, National Cheng Kung University, 138 Sheng-Li Road, Tainan 704, Taiwan (China)
  • 2. Department of Microbiology and Immunology, National Defense Medical Center, Taipei 114, Taiwan (China)
  • 3. Division of Method Development and Analysis, Institute of Occupational Safety and Health, Shijr City 221, Taipei, Taiwan (China)

Description

The aim of this study is to validate isotope-dilution electrospray ionization tandem mass spectrometry (ESI-MS-MS) method with a dual-loop cleanup device for simultaneous quantitation of two benzene metabolites, trans, trans-muconic acid (ttMA) and S-phenylmercapturic acid (SPMA), in human urine. In this study, a pooled blank urine matrix from rural residents was adopted for validation of the analytical method. The calibration curve, detection limit, recovery, precision, accuracy and the stability of sample storage for the system have been characterized. Calibration plots of ttMA and SPMA standards spiked into two kinds of urine matrixes over a wide concentration range, 1/32-8-fold biological exposure indices (BEIs) values, showed good linearity (R > 0.9992). The detection limits in pooled urine matrix for ttMA and SPMA were 1.27 and 0.042 μg g-1 creatinine, respectively. For both of ttMA and SPMA, the intra- and inter-day precision values were considered acceptable well below 25% at the various spiked concentrations. The intra- and inter-day apparent recovery values were also considered acceptable (apparent recovery >90%). The ttMA accuracy was estimated by urinary standard reference material (SRM). The accuracy reported in terms of relative error (RE) was 5.0 ± 2.0% (n = 3). The stability of sample storage at 4 or -20 deg. C were assessed. Urinary ttMA and SPMA were found to be stable for at least 8 weeks when stored at 4 or -20 deg. C. In addition, urine samples from different benzene exposure groups were collected and measured in this system. Without tedious manual sample preparation procedure, the analytical system was able to quantify simultaneously ttMA and SPMA in less than 20 min

Additional details

Identifiers

DOI
10.1016/j.aca.2005.08.075;
PII
S0003-2670(05)01478-9;

Publishing Information

Journal Title
Analytica Chimica Acta
Journal Volume
555
Journal Issue
1
Journal Page Range
p. 34-40
ISSN
0003-2670
CODEN
ACACAM

Optional Information

Copyright