Published December 2016 | Version v1
Journal article

High-resolution monochromator for iron nuclear resonance vibrational spectroscopy of biological samples

  • 1. Japan Synchrotron Radiation Research Institute (JASRI/SPring-8), Sayo, Hyogo (Japan)
  • 2. Department of Chemistry, University of California, Davis, CA (United States)
  • 3. Kyoto University, Research Reactor Institute, Kumatori, Osaka (Japan)

Description

A new high-resolution monochromator for 14.4-keV X-rays has been designed and developed for the Fe nuclear resonance vibrational spectroscopy of biological samples. In addition to high resolution, higher flux and stability are especially important for measuring biological samples, because of the very weak signals produced due to the low concentrations of Fe-57. A 24% increase in flux while maintaining a high resolution better than 0.9 meV is achieved in the calculation by adopting an asymmetric reflection of Ge, which is used as the first crystal of the three-bounce high-resolution monochromator. A 20% increase of the exit beam size is acceptable to our biological applications. The higher throughput of the new design has been experimentally verified. A fine rotation mechanics that combines a weak-link hinge with a piezoelectric actuator was used for controlling the photon energy of the monochromatic beam. The resulting stability is sufficient to preserve the intrinsic resolution. (author)

Additional details

Identifiers

Publishing Information

Journal Title
Japanese Journal of Applied Physics
Journal Volume
55
Journal Issue
12
Journal Page Range
p. 122401.1-122401.4
ISSN
0021-4922

Optional Information

Notes
20 refs., 5 figs., 1 tab.