Published November 2018 | Version v1
Journal article

Radiation exposure of the investigator during navigated fusion of 124iodine PET imaging and ultrasound

  • 1. Clinic of Nuclear Medicine, Jena University Hospital, Jena (Germany)

Description

To assess the radiation exposure of the investigator during navigated fusion of nuclear medicine images with ultrasound after application of I-124. Dosimetry with two different types of thermoluminescent detectors (TLD) was performed in 25 patients. The dose rate at the patient's neck was measured with a calibrated dose rate meter (DRM) and served as the standard of reference. The average exposure per investigation at the patient's neck measured by LiF:Mg,Cu,P TLDs (cumulative: 212 μSv), LiF:Mg,Ti TLDs (cumulative: 112 μSv) and DRM (cumulative: 150.3 μSv). The radiation exposure of the hand during navigated fusion of nuclear medicine imaging with 124I and ultrasound with a mean duration of 13 min is low and comparable between different methods. Yearly examinations are not expected to add a relevant cumulative risk. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncy036

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
181
Journal Issue
4
Journal Page Range
p. 368-373
ISSN
0144-8420

Optional Information

Notes
22 refs.