Description

In this work it is discussed the technical conditions of the implantation of radiation qualities adequate for dosemeter applied to mammography at the Radiological Science Laboratory (RSL) of the University of State of Rio de Janeiro, Brazil. It performed tests for evaluation the homogeneity of the radiation field, the scattering, the half-layers and the system stability. The test results point for adequacy of the laboratory installed at the LCR for calibration of the mentioned dosemeters in seven qualities of radiation and the expanded uncertainty with better capacity of measurement at those qualities was of ±1.8% (k = 2)

Files

42035274.pdf

Files (194.4 kB)

Name Size Download all
md5:1a18d54298cf4927c0f3b36c275333f5
194.4 kB Preview Download

Additional details

Additional titles

Original title (Portuguese)
Implantacao das qualidades de radiacao para mamografia, segundo os criterios da IEC 61627 e do TRS 457

Publishing Information

Imprint Pagination
7 p.
Report number
INIS-BR--8695

Conference

Title
5. Brazilian congress on metrology
Original Conference Title
5. Congresso brasileiro de metrologia. Metrologia 2009
Acronym
Metrology 2009
Dates
9-13 Nov 2009
Place
Salvador, BA (Brazil)

Optional Information

Notes
Code: ID390.pdf