Preliminary commissioning investigations with the DMOS-RPC optical-CT Scanner
Creators
- 1. Department of Radiation Oncology, Duke University Medical Center, Durham, NC (United States)
Description
A midsized broad beam Optical-CT scanner is being developed for collaborative research between Duke and the Radiological Physics Center (RPC). The Duke Midsized Optical-CT Scanner (DMOS-RPC) is designed to be compatible with several of the RPC phantoms, including the head and neck, stereotactic SRS, and lung phantoms. Preliminary data investigating the basic performance of the scanner is described. Two 10 cm PRESAGE cylinders were irradiated with simple test plans. Projections of ∼80 μm resolution of each dosimeter were collected at 1 degree intervals over a full 360 degrees both before and after irradiation. 3 dimensional reconstructions of attenuation coefficients throughout the dosimeter were computed with 1 mm3 resolution. Scans were normalized to the calculated dose distribution and a 3D comparison was made with a commissioned treatment planning system. Initial results indicate DMOS-RPC can produce accurate relative dose distributions with high spatial resolution (up to 1 mm3 in 3D) in less than 30 minutes (acquisition and reconstruction). A maximum dose of ∼3.6Gy was delivered in these tests, and observed noise was ∼2% for 1 mm3 reconstructions. Good agreement is observed with the planning system in these simple distributions, indicating promising potential for this scanner.
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/250/1/012078Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 250
- Journal Issue
- 1
- Journal Page Range
- [4 p.]
- ISSN
- 1742-6596
Conference
- Title
- 6. international conference on 3D radiation dosimetry
- Acronym
- IC3DDose
- Dates
- 22-26 Aug 2010
- Place
- Hilton Head Island, SC (United States)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 42053593
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ATTENUATION; BEAMS; COMMISSIONING; COMPARATIVE EVALUATIONS; COMPUTERIZED SIMULATION; COMPUTERIZED TOMOGRAPHY; DOSEMETERS; HEAD; IRRADIATION; LUNGS; NECK; NOISE; PHANTOMS; PLANNING; RADIATION DOSE DISTRIBUTIONS; SPATIAL RESOLUTION; THREE-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- BODY; DIAGNOSTIC TECHNIQUES; EVALUATION; MEASURING INSTRUMENTS; MOCKUP; ORGANS; RESOLUTION; RESPIRATORY SYSTEM; SIMULATION; STRUCTURAL MODELS; TOMOGRAPHY