Published September 12, 1986 | Version v1
Journal article

Testing the accuracy of neutron activation analysis of silver in biological materials

  • 1. Reading Univ. (UK). Dept. of Chemistry

Description

While silver can be efficiently collected from biological materials into lead beads by conventional fire assay, reagent blanks make this method unsuitable for determining the metal at concentrations less than 1 ppm. A radiochemical neutron activation analysis method is described which avoids this problem and can give reliable results down to 1 ng g-1. The method was tested using various reference materials (RM), and it is concluded that NIES Mussel and Pepperbush are the best characterized RMs for determining traces of silver. (author)

Additional details

Publishing Information

Journal Title
J. Radioanal. Nucl. Chem.
Journal Volume
106
Journal Issue
4
Series
J. Radioanal. Nucl. Chem. ;Letters.
Journal Page Range
193-203
ISSN
0236-5731
CODEN
JRNCD

Optional Information

Notes
35 refs.; 3 tables.