Published September 12, 1986
| Version v1
Journal article
Testing the accuracy of neutron activation analysis of silver in biological materials
Description
While silver can be efficiently collected from biological materials into lead beads by conventional fire assay, reagent blanks make this method unsuitable for determining the metal at concentrations less than 1 ppm. A radiochemical neutron activation analysis method is described which avoids this problem and can give reliable results down to 1 ng g-1. The method was tested using various reference materials (RM), and it is concluded that NIES Mussel and Pepperbush are the best characterized RMs for determining traces of silver. (author)
Additional details
Publishing Information
- Journal Title
- J. Radioanal. Nucl. Chem.
- Journal Volume
- 106
- Journal Issue
- 4
- Series
- J. Radioanal. Nucl. Chem. ;Letters.
- Journal Page Range
- 193-203
- ISSN
- 0236-5731
- CODEN
- JRNCD
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 18034871
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- ACCURACY; BIOLOGICAL MATERIALS; CALIBRATION STANDARDS; COMPARATIVE EVALUATIONS; DRY ASHING; EXPERIMENTAL DATA; FLUIDIZED-BED COMBUSTION; LEAD; NEUTRON ACTIVATION ANALYSIS; SAMPLE PREPARATION; SILVER; TRACE AMOUNTS
- Descriptors DEC
- ACTIVATION ANALYSIS; CHEMICAL ANALYSIS; CHEMICAL REACTIONS; COMBUSTION; DATA; ELEMENTS; EVALUATION; INFORMATION; MATERIALS; METALS; NONDESTRUCTIVE ANALYSIS; NUMERICAL DATA; OXIDATION; TRANSITION ELEMENTS
Optional Information
- Notes
- 35 refs.; 3 tables.