Published June 2001 | Version v1
Journal article

Measurement of the scattered spatial dose distribution for the mobile X-ray radiography

  • 1. Taegu Health college, Taegu (Korea, Republic of)
  • 2. Yeungnam Univ, Kyeongsan (Korea, Republic of)

Description

The spatial distribution of the scattered dose for mobile X-ray radiography is measured. The scattered X-ray exposures at the radius of 50, 100 and 150 cm from the irradiation center are 880, 180 and 50 μR, respectively. This scattered X-rays can be reduced to 60% by inserting the portable shield made by 0.4 mm copper sheet sandwiched in two plywoods

Additional details

Publishing Information

Journal Title
Journal of the Korean Society of Radiological Technology
Journal Volume
24
Journal Issue
1
Series
7 refs, 10 figs
Journal Page Range
p. 23-26
ISSN
1226-2854

INIS

Country of Publication
Korea, Republic of
Country of Input or Organization
Korea, Republic of
INIS RN
32067182
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
PATIENTS; RADIOLOGY; SHIELDS; SPATIAL DOSE DISTRIBUTIONS; X-RAY RADIOGRAPHY
Descriptors DEC