Published 1989 | Version v1
Report Open

Spatial resolution of electromagnetic liquid krypton calorimeter

  • 1. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Serpukhov (USSR). Inst. Fiziki Vysokikh Ehnergij
  • 2. Moskovskij Fiziko-Tekhnicheskij Inst., Moscow (USSR)

Description

The influence of induction effect and multiple Coulomb scattering on spatial resolution of electromagnetic liquid krypton calorimeter, when determining gamma-quanta coordinate by the charge division in the capacitance line, is studied. Monte Carlo calculation confirmed the assumption, that for l=10mm strip width detector the induction effect must not influence significantly the finite results. With corresponding values of coupling constant and strip width the detector spatial resolution is defined as ∼ l/√12 mm. When determining gamma-quantum coordinate by charge division in the capacitance line for E=0.1-1GeV the strip decrease to less than 3mm is meaningless due to contribution of multiple Coulomb scattering, which defines spatial resolution limit as ∼ 0.6-0.7 mm. 11 refs.; 5 figs.; 2 tabs

Availability note (English)

MF available from INIS under the Report Number.

Files

21080330.pdf

Files (449.3 kB)

Name Size Download all
md5:7f69bada9c3590e8438ddbd6854065b7
449.3 kB Preview Download

Additional details

Additional titles

Subtitle (English)
Monte-Carlo simulation
Original title (Russian)
Koordinatnoe razreshenie ehlektromagnitnogo zhidkokriptonnogo kalorimetra
Original subtitle (Russian)
Modelirovanie metodom Monte-Karlo.

Publishing Information

Imprint Pagination
21 p.
Report number
IFVE-ONF--89-135

Optional Information

Notes
Submitted to NIM.