Published October 1976
| Version v1
Report
Open
A calibrated 6Li glass detector for fast neutrons. Description and use
Description
A fast neutron detector which efficiency has been carefully measured is described. The time response of this detector is better than 1.2 nanosecond. This detector is used with fast-time-of-flight technique. Its efficiency is known within 2.5% between 5 and 180keV and within 5% between 300 and 1200keV
Availability note (English)
MF available from INIS under the Report Number.Abstract (French)
On decrit un detecteur de neutrons rapides dont on a mesure soigneusement l'efficacite. Ce detecteur possede une resolution en temps superieure a 1,2 nanoseconde et est utilise avec la technique du temps de vol rapide. Son efficacite est connue a 2,5% pres entre 5 et 180keV et a 5% entre 300 et 1200keVFiles
8311140.pdf
Files
(1.5 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:917e2eb5cfcde16bd9f0f2ba3417972a
|
1.5 MB | Preview Download |
Additional details
Additional titles
- Original title (French)
- Detecteur etalonne de neutrons rapides constitue par un verre au
Publishing Information
- Imprint Pagination
- 38 p.
- Report number
- CEA-N--1904
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 8311140
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- DOPED MATERIALS; EFFICIENCY; FAST NEUTRONS; GLASS SCINTILLATORS; KEV RANGE 10-100; KEV RANGE 100-1000; LITHIUM ADDITIONS; LITHIUM 6; NEUTRON DETECTORS; TIME-OF-FLIGHT METHOD
- Descriptors DEC
- BARYONS; ELEMENTARY PARTICLES; ENERGY RANGE; FERMIONS; HADRONS; ISOTOPES; KEV RANGE; LIGHT NUCLEI; LITHIUM ISOTOPES; MEASURING INSTRUMENTS; NEUTRONS; NUCLEI; NUCLEONS; ODD-ODD NUCLEI; PHOSPHORS; RADIATION DETECTORS; STABLE ISOTOPES