Published November 1987
| Version v1
Journal article
Fabrication and study on the cylindrical scintillation counters of the ARES spectrometer
Creators
- 1. Joint Inst. for Nuclear Research, Dubna (USSR)
- 2. AN Belorusskoj SSR, Minsk. Inst. Fiziki
- 3. Tbilisskij Gosudarstvennyj Univ. (USSR)
Description
Fabrication technique for plastic scintillation counters, which are the elements of the cylindrical 673-mm diameter hodoscope, is described. The plastic scintillation counter consists of a scintillator, made on the polystyrene base and with 600x87x5 mm dimensions; transient and base lightdiodes, made from polymethylmethacrylate, and photomultiplier. The optical contact between the base lightdiode and photocathode is carried out by vaseline. The investigation results of counter amplitude and time characteristics, which were conducted by means of 90Sr β-source. At β-source mounting in the scintillator centre the counter amplitude resolution constitutes 80%, and its time resolution equals to 1 ns
Additional details
Additional titles
- Original title (Russian)
- Изготовление и исследование цилиндрических сцинтилляционных счетчиков спектрометра ARES
Publishing Information
- Journal Title
- Prib. Tekh. Ehksp.
- Journal Issue
- no.6
- Series
- Prib. Tekh. Ehksp.
- ISSN
- 0032-8162
- CODEN
- PRTEA
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- USSR
- INIS RN
- 19073318
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CYLINDRICAL CONFIGURATION; HODOSCOPES; LIGHT PIPES; PERFORMANCE TESTING; PLASTIC SCINTILLATION DETECTOR; PRESSING; TIME RESOLUTION
- Descriptors DEC
- CONFIGURATION; FABRICATION; MATERIALS WORKING; MEASURING INSTRUMENTS; RADIATION DETECTORS; RESOLUTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; TESTING; TIMING PROPERTIES