Published 1987 | Version v1
Report Open

JINR phasotron beam monitors

Description

JINR phasotron beam monitoring system permitting to realize continuous control of 680 MeV, 10 μA, accelerated proton beam parameters is described. A scintillation NaI γ-detector is the main element of the monitor for determination of the radius losses. The detector is placed at the distance of 10 m from the chamber. Radial position of the beam is defined by means of a gauge. Five pairs of magnetoinduction sensors are installed in the radius range from 107 to 262 cm with an equal step. Secondary-emission detectors of two types: multielectrode and differential are used. The monitoring system utilizes computerized means of the on-line control system of the phasotron

Files

19106133.pdf

Files (1.1 MB)

Name Size Download all
md5:d92bc35539b0f113c6c7b4e366510966
1.1 MB Preview Download

System files (14.5 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Устройства диагностики пучка фазотрона OIYaI

Publishing Information

Imprint Title
Proceedings of the 10. All-union conference on accelerators of charged particles. Vol. 1
Journal Page Range
p. 56-60.
Report number
JINR-D--9-87-105(v.1)

Conference

Title
10. All-union conference on charged-particle accelerators.
Dates
21-23 Oct 1986.
Place
Dubna (USSR).

Optional Information

Notes
3 refs.; 10 figs. Imprint:Trudy 10. Vsesoyuznogo soveshchaniya po uskoritelyam zaryazhennykh chastits. Tom 1.